Maverick qPCR System (MQ4 Series) Instruction Manual (V1.0) - Chapter 6

Prev Article:Chapter 5 Instrument Maintenance


The MQ4 model is an ultra-portable qPCR system with a large 10-inch touch screen for integrated operation. The product is developed based on our multi-channel fluorescence imaging optical system with CMOS bio-image sensor. The product can be widely used in universities and research institutes, CDC, Entry-Exit Inspection and Quarantine Bureau, Public Security Criminal Evidence Identification Center, veterinary stations, food companies and pharmaceutical companies.

Chapter 6 FAQ

No.

Failure Phenomenon

Cause Analysis

Processing

1

The screen   shows a black screen

Screen motherboard damage

Must replace the screen motherboard, please contact with the supplier   or manufacturer

If the screen lock function is incorrectly operated, the screen enters   the black screen standby mode

Close the lock screen, and enter the test interface directly after the   startup

Screen damage

Need to replace the screen, please contact the supplier or   manufacturer

Software   prompts "Abnormal auxiliary heating

Auxiliary temperature selftest abnormal

Please restart the instrument first to confirm, if still can not solve   please contact the supplier or manufacturer

Abnormal   heat-up curve of hot cover

Power supply problems

Verify that the power is plugged in properly

Hot cover assembly problems

Please restart the instrument first to confirm, if still can not solve   please contact the supplier or manufacturer

2

Unable to turn on

Power supply problems

Verify that the power is plugged in properly

Switch or   power cable damage

Please restart the instrument first to confirm, if still can not solve   please contact the supplier or manufacturer

3

USB flash   drive export failed

The USB disk   is not in good contact

Reinsert the USB flash drive for confirmation

Can try to use WIFI, Bluetooth export function

If the problem persists, contact the supplier or manufacturer

4

HID connection exception

Screen communication exception

Please restart the instrument first to confirm, if still can not solve   please contact the supplier or manufacturer

5

No experimental data after instrument operation

Incorrect setting of experimental parameters

Verify that the thermal cycling parameters and sample parameters are   set correctly, and set the fluorescent markers

6

Report upload and print function exception

WIFI setting error

Please check if the instrument WIFI setting is correct, please connect   to the network correctly

7

Test time and report time are not synchronized

Time synchronization without network connection

The default time is used   when the device is not connected to the network. Please connect to the   network for immediate time update

8

Software prompts "Please close the hot cover"

Instrument flip cover is not closed in place

Re-close the flap to ensure it is closed in place.


Professional Portable qPCR System Manufacturer - Anitoa

Anitoa is a professional R&D and manufacturing company of molecular diagnostic qPCR instruments and reagents, with independent intellectual property rights of core chip technology, optical technology, rapid heating, microfluidic control and supporting reagents and other patented technologies, so that the instruments developed by Anitoa are characterized by quick and easy small instruments, allowing the development of expensive and complex large PCR instruments into truly portable POCT products.